FAQ
Digi Scope DS601AX
MULTI-FORMAT & MULTI STANDARD ON SCREEN DISPLAY ANALYZER MEASUREMENT AND MONITORING UNIT.
Send information to a colleague
 
Can I use a PC to download a screen shot?

What is the difference between the AX and BX?

I only get one analogue audio bar on screen

Is there a reset for the instrument?

I recall a user set-up and the instrument locks up

In parade mode I only get Luminance (white), R and B traces no G

I have connected an active SDI signal but the instrument does not identify the signal as being present

What does the jitter figure on the alarm screen represent

What is its weight

What is its dimensions

The instrument will not start

The instrument reset does not work completely

The output monitor displays a green screen

Real time clock has lost its setting

When turned on instrument did not return to last powered down state

 

Can I use a PC to download a screen shot?

Yes download the RS232 program from the service/support area

 

What is the difference between the AX and BX

The BX has reduced analyzer functionality, no Cursor Measurement, Bow Tie, Error count, Download, Logger or Safe Area features.

 

I only get one analogue audio bar on screen

Power down the instrument, hold the 'Y' button in then power on again (you are resetting the instrument)

 

Is there a reset for the instrument

Power down the instrument, hold the 'Y' button in then power on again (unless you have a very early instrument then you need to hold the configure button down)

 

I recall a user set-up and the instrument locks up

This is due to the user setup you selected having nothing in it, to solve the problem turn the instrument off, hold the 'B' blue button down and power on .(this preloads all the user set-ups with a simple factory recall set-up)

 

In parade mode I only get Luminance (white), R and B traces no G

Select configure then set-up then change the component output from 3 channel to 4 channel output

 

I have connected an active SDI signal but the instrument does not identify the signal as being present

On the back panel there is an SDI input and an SDI loop out, usually the cause is that the signal is on the loop out not hte input, there is a buffer between the in and out so it is not reversible

 

The instrument will not start?

there is an internal battery that if it falls below about 1.5 volts will not allow the instrument to start This should only occur on rare occasions for instruments over 3 years old.

contact your local distributor to arrage for a new battery to be fitted

 

The instrument reset does not work completely?

there is an internal battery that if it falls below about 3.0 volts will not allow the instrument to fully function. This should only occur on rare occasions for instruments over 3 years old.

contact your local distributor to arrage for a new battery to be fitted

 

The output monitor displays a green screen?

there is an internal battery that if it falls below about 3.0 volts will not allow the instrument to fully function. This should only occur on rare occasions for instruments over 3 years old.

contact your local distributor to arrage for a new battery to be fitted

 

Real time clock has lost its setting ?

there is an internal battery that if it falls below about 3.0 volts will not allow the instrument to fully function. This should only occur on rare occasions for instruments over 3 years old.

contact your local distributor to arrage for a new battery to be fitted

 

When turned on instrument did not return to last powered down state?

there is an internal battery that if it falls below about 3.0 volts will not allow the instrument to fully function. This should only occur on rare occasions for instruments over 3 years old.

contact your local distributor to arrage for a new battery to be fitted

 

What does the jitter figure on the alarm screen represent

The figure shown directly relates to how hard the PLL in the instrument is working to lock on to the input SDI signal 0 is the best figure where the PLL is having no difficulty locking a high figure of about 40 is a serious problem for stability and close to the PLL losing lock

 

What is its weight

 

3 kilo

 

What is its dimensions

in cm

Length:- 32

Width:- 23

Height:- 5

    